Rapid/Easy/Accurate TOC Measurement
- Low concentration stability (quantification limit: 4 ppb) - Operates immediately after power is turned on - No carrier gas required - Calibration frequency is once a year - Short measurement time (4 measurements per sample in 14 minutes) - Compact design - Three types: lab type / portable type / online type - Inorganic carbon removal device (ICR) built-in as standard - Optional auto-sampler for automatic measurement of up to 63 samples (40 mL vials) - Custom multi-point calibration protocol compliant with notification methods for the water sector
Inquire About This Product
basic information
- Total Organic Carbon (TOC) Measurement Principle: Wet UV oxidation + gas permeable membrane conductivity measurement method Measurement Range: 4 ppb to 50 ppm TOC Accuracy: ±2% or 0.5 ppb, whichever is greater Reproducibility: RSD less than 1% Display Digits: 3 digits Calibration Frequency: Recommended every 12 months Measurement Interval: 2 minutes Sample Water Temperature: 5 to 95°C Sample Water Pressure: Up to 100 psig Sample Water Flow Rate: 50 mL/min or more Sample Flow Rate: 0.5 mL/min Interfering Substances: Not affected by coexisting ions other than organic substances - Equipment Power Requirements: 100 to 240 V, 50/60 Hz, 100 VA Ambient Temperature: 5 to 40°C Humidity: Up to 95% without condensation Dimensions: Varies by model Weight: Varies by model (9.8 kg to 16.2 kg) *For detailed specifications, please refer to the catalog.
Price range
Delivery Time
Applications/Examples of results
Measurement of tap water Measurement of environmental water Cost reduction (coagulants / activated carbon / ion exchange resin) Determination of optimal water treatment processes according to seasonal variations of raw water Detection of chemical leaks
Detailed information
-
- Sievers' unique gas permeable membrane conductivity measurement method. In the TOC measurement method for conductivity, the conductivity indicated by the carbon dioxide (carbonate ions) produced after the oxidation and decomposition of organic matter is measured to calculate the carbon concentration. The Sievers M500e enables accurate TOC measurement by using a "gas permeable membrane module," which separates carbon dioxide from the sample water to the pure water line within the analyzer through a thin Teflon film, thereby eliminating the influence of other ions.
Line up(4)
Model number | overview |
---|---|
SPRD77120-01 | TOC Analyzer M5310C Laboratory Type <Built-in ICR> |
SPRD77220-01 | TOC Analyzer M5310C Portable Type <Built-in ICR> |
SPRD77320-01 | TOC Analyzer M5310C Online Type <Built-in ICR> |
SPRD77420-01 | TOC Analyzer M5310C Online Type <Built-in ICR / Dual Channel Model> |
Company information
Since our establishment in 1967, we have been committed to the development and dissemination of water quality measuring instruments that anyone can use quickly, easily, and with high precision. In today's world, where effective utilization of the limited resource of water is required alongside environmental conservation, we aim to be a total solution provider that responds to all water environments by leveraging our water quality measuring instruments, technology, and services. Additionally, we are dedicated to utilizing the latest technology and our years of expertise to deliver user-friendly and reliable water quality measuring instruments more affordably, quickly, and widely.