Continuous monitoring of ultra-low concentration TOC in ultrapure water for the semiconductor and electronics industries! Rapid measurement across a wide range!
The Sievers M9e can accommodate a variety of applications due to its unique gas-permeable membrane conductivity measurement method and wide measurement range (0.03 ppb to 50 ppm). The optional turbo mode (analysis time of 4 seconds) is useful for monitoring recovered water. Features of the Sievers M9e: - Matching of measurement deviations - Stability in ultra-low concentration ranges - Automatic operation (calibration / verification / data analysis, etc.) - 12 months of calibration stability - Three models to choose from based on application (laboratory type / portable type / online type) - Sample conductivity measurement function (optional) allows simultaneous measurement of TOC and conductivity - Auto-sampler (optional) can automatically analyze up to 63 samples (40 mL each).
Inquire About This Product
basic information
- Common Specifications (Normal Mode & Turbo Mode) Measurement Principle: Wet UV oxidation + Gas permeable membrane conductivity measurement method Sample Water Flow Rate: 50 mL/min or more Sample Temperature: 5–95°C Sample Water Pressure: Up to 100 psig Interfering Substances: Not affected by coexisting ions other than organic matter Calibration Stability: Stable for normally 12 months Display Digits: 3 digits - Normal Mode Measurement Range: 0.03 ppb to 50 ppm TOC Accuracy: ± 2% or ±0.5 ppb, whichever is larger Reproducibility: RSD less than 1% Measurement Time: 2 minutes Flow Rate Inside Analyzer: 0.5 mL/min - Turbo Mode Measurement Range: 0.2 ppb to 10 ppm TOC Accuracy: Within ± 3% Reproducibility: RSD less than 2% Measurement Time: 4 seconds Flow Rate Inside Analyzer: 1.1 mL/min
Price range
Delivery Time
Applications/Examples of results
Ultra-pure water production equipment for the semiconductor industry: raw water/RO water/recovery water Equipment cleaning evaluation Inorganic acid purity management Plating solution contamination evaluation Water treatment membrane performance evaluation Cooling water/boiler water leak detection
Detailed information
-
- Sievers' unique gas permeable membrane conductivity measurement method In the TOC measurement method for conductivity, the conductivity indicated by the carbon dioxide (carbonate ions) produced after the oxidation and decomposition of organic matter is measured to calculate the carbon concentration. The Sievers M500e enables accurate TOC measurement by using a "gas permeable membrane module," which separates carbon dioxide from the sample water to the pure water line within the analyzer through a thin Teflon film, thus eliminating the influence of other ions.
Line up(3)
Model number | overview |
---|---|
SPRD77130-01 | TOC Meter M9e Laboratory Type |
SPRD77230-01 | TOC Meter M9e Portable Type |
SPRD77330-01 | TOC Meter M9e Online Type |
Company information
Since our establishment in 1967, we have been committed to the development and dissemination of water quality measuring instruments that anyone can use quickly, easily, and with high precision. In today's world, where effective utilization of the limited resource of water is required alongside environmental conservation, we aim to be a total solution provider that responds to all water environments by leveraging our water quality measuring instruments, technology, and services. Additionally, we are dedicated to utilizing the latest technology and our years of expertise to deliver user-friendly and reliable water quality measuring instruments more affordably, quickly, and widely.