For the verification of the SF double-shell tank test.
- Easily detect the adhesion and gap condition of the detection layer inside the shell wall of the SF double-shell tank from the surface of the outer shell plate (FRP). - Measure the thickness of the FRP using ultrasound.
Inquire About This Product
basic information
Measurement method: Pulse reflection method (single probe method) Operating frequency: 1–2 MHz Display method: LCD digital (with EL backlight) Measurement range: 1.0–50.0 mm Mode automatic switching (standard/thick plate) Measurement accuracy: ±0.1 mm Sound speed adjustment range: 500–9999 m/sec Probe: 2C20N-DL type Power supply: AA alkaline battery (1.5V) x1 Approximately 12 hours (about 8 hours with EL lighting) Operating temperature range: -10 to 50°C (no condensation) Dimensions and weight: 66 (W) × 180 (H) × 28 (D) mm, 360 g
Price range
Delivery Time
Applications/Examples of results
For the verification of the SF double-shell tank test.
Company information
Sankoh Electronic Research Institute's measuring and inspection equipment strongly supports quality control operations with numerous advantages, including: ◎ High precision and high reliability ◎ Hardware that withstands rigorous professional use ◎ User-centric design philosophy prioritizing operability ◎ Comprehensive support system backed by extensive experience and know-how.