Method for expanding the dynamic range of a spectrum analyzer equipped with a tracking generator.
[Application Example] Method for Expanding the Dynamic Range of Spectrum Analyzers Equipped with Tracking Generators (TG) Using Low-Noise Amplifiers
This introduces a method for expanding the dynamic range of spectrum analyzers equipped with a tracking generator (TG) by using a low-noise amplifier. 【Product Introduction】 ■ Signal Analyzer MSA538TG ■ Spectrum Analyzer MSA338TG ■ Spectrum Analyzer MSA438TG ■ Low Noise Amplifier MAP302 *For more details, please refer to the PDF document or feel free to contact us.
Inquire About This Product
basic information
[Application] When measuring using the TG function with a TG-equipped spectrum analyzer, the dynamic range is limited by the TG output level. Therefore, in measurements that require a high TG output level, such as radio wave shielding performance tests, there may be insufficient dynamic range. [Solution] By utilizing a low-noise amplifier, the lower limit of the dynamic range when measuring with the TG function can be expanded. This is an example of expanding the dynamic range of the TG-equipped 3.3GHz signal analyzer (real-time spectrum analyzer) MSA538TG using the low-noise amplifier MAP302. [System Configuration] ■ Signal Analyzer MSA538TG ■ Spectrum Analyzer MSA338TG ■ Spectrum Analyzer MSA438TG ■ Low Noise Amplifier MAP302 *For more details, please refer to the PDF document or feel free to contact us.
Price range
Delivery Time
Applications/Examples of results
For more details, please refer to the PDF document or feel free to contact us.
catalog(1)
Download All CatalogsCompany information
● An independent, domestically produced advanced development company in the high-frequency (RF) field ● One-stop development, manufacturing, sales, and service business for electronic measuring instruments, information and communication devices, and environmental-related equipment ● A worldwide sales network with 18 domestic agents and 19 overseas agents