Oxygen Monitor OXYMAN Series - Inspection and Calibration Information

1. Daily Inspections by Users
1) Operational checks at the start of each work shift (displaying around 20.9% and stable)
If the displayed value is unstable, it indicates sensor degradation, and sensor replacement is necessary. *
2) Atmospheric calibration using the "CAL" button approximately once a month (displaying 20.9%)
Even within a month, if the displayed value deviates below 20.4 or above 21.4, please perform this atmospheric calibration.
This operation updates the sensor's remaining capacity indicator.
When it reaches 1 block remaining, please promptly arrange for sensor replacement. *
* Please contact us at the contact information below.
2. Inspections and Calibration by the Manufacturer
1) For purposes other than oxygen deficiency monitoring, the above 1. applies.
2) In cases where the oxygen monitor is used continuously for oxygen deficiency monitoring and malfunctions are treated as accidents, in addition to the above 1., we recommend planned regular maintenance (once a year).
If there are stricter regulations regarding maintenance of your company's safety devices, please follow those.