Oxygen Monitor OXYMAN Series: How to Replace the Separated Oxygen Sensor

Thank you very much for your continued support of our products.
Our oxygen monitor OXYMAN series (OXYMAN, OXYMAN Plus) includes both built-in and separate types.
For the built-in type, the method for replacing the oxygen sensor is as follows:
<Send the entire oxygen monitor unit from the customer to us
→ We will replace the built-in sensor and perform span adjustment
→ We will return it to the customer>
In contrast, for the separate type, in addition to the above method, we also accept orders for the separate sensor unit (oxygen sensor, case, and sensor cable unit) individually. (In this case, the replacement will be done on the customer’s side.)
When placing an order, please be sure to include the serial number of the oxygen monitor unit to which the sensor unit will be connected (as indicated on the label on the back of the unit).
Additionally, if a considerable amount of time has passed since the purchase of the oxygen monitor unit (more than 8 years) or if there are issues with the oxygen monitor unit itself, we recommend following the same method as for the built-in oxygen sensor.
TEL 0296-74-4631 FAX 0296-74-4780
E-mail: k_hanbai@taiei-eng.co.jp

Inquiry about this news
Contact Us OnlineMore Details & Registration
Details & Registration
Related Links
Taiei Engineering